Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Intercomparison of the Realization of the ITS-90 at the Freezing Points of Al and Ag among European NMIs (CROSBI ID 137833)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Heyer, D. ; Noatsch, U. ; Tegeler, E. ; Anagnostou, M. ; Turzo-Andras, E. ; Antonsen, I. ; Augevicius, V. ; Bojkovski, J. ; Bronnum, A. ; Chimenti, V. et al. Intercomparison of the Realization of the ITS-90 at the Freezing Points of Al and Ag among European NMIs // International journal of thermophysics, 28 (2008), 6; 1964-1975. doi: 10.1007/s10765-007-0283-y

Podaci o odgovornosti

Heyer, D. ; Noatsch, U. ; Tegeler, E. ; Anagnostou, M. ; Turzo-Andras, E. ; Antonsen, I. ; Augevicius, V. ; Bojkovski, J. ; Bronnum, A. ; Chimenti, V. ; Duris, V. ; Filipe, E. ; Gaita, S. ; Gray, J. ; Head, D. ; Grudniewicz, E. ; Ivarsson, J. ; Kalemci, M. ; Kerkhof, O. ; Lobo, I. ; Nemeth, S. ; Pokhodun, A. ; Ranostaj, J. ; Renaot, E. ; Rosenkranz, P. ; Smid, M. ; Steur, P. ; Steiner, A. ; Valin, M. ; Veliki, Tomislav ; Weckström, T.

engleski

Intercomparison of the Realization of the ITS-90 at the Freezing Points of Al and Ag among European NMIs

The EUROMET.T-K4 comparison is the regional extension of CCT-K4, an intercomparison of the realizations of the freezing points of Al (660.323°C) and Ag (961.78°C). The intercomparison was organized in four loops. Long-stem standard platinum resistance thermometers (SPRTs) were used as traveling standards: 25 Ω thermometers to be used only at the Al freezing point and two high-temperature standard platinum resistance thermometers (HTSPRTs) to be used only at the Ag freezing point in each loop. Parallel to the measurements with thermometers, the pilots and sub-pilots organized an internal intercomparison using an Ag fixed-point cell. Most HTSPRTs showed a strong drift which is mainly due to mechanical stress and poisoning of the sensor by impurities. This drift can be partially compensated by a correction based on Matthiessen’ s rule. An evaluation of the data taking into account both HTSPRTs in each loop, the linkage of the sub-pilots by measurements at the Ag freezing point, and a possible compensation according to Matthiessens’ s rule, allows calculation of the results of the participants’ measurements at the Ag freezing point. The results of the participating laboratories are summarized, and proposals for key comparison reference values and linking of the results to CCT-K3 and CCT-K4 are presented.

Al freezing point; CCT; EUAl freezing point; Ag ROMET; Fixed points; High-temperature standard platinum resistance thermometers; Key comparison; Matthiessen’ s rule