A custom designed system to measure the corticospinal tract jitter (CROSBI ID 81313)
Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Mihelin, Marjan ; Liščić, Rajka M.
engleski
A custom designed system to measure the corticospinal tract jitter
Typical latency of an individual limb muscle response to magnetic or electric stimulation of the human cortex is in the range of 10-50 ms. For the latency varaiability, i.e., jitter studies, a resulution of at least 20 us is needed. Commercially available EMG equipment needs custom-designed upgrading to allow for such studies. Two solutions were designed: (i) a hardware unit allowing an adjustable delay of data acquisition after the delivered stimuli; and (ii) diverting of the amplified biological signal and the EMG equipment trigger to an external computer equipped with an analogue-to-digital conversion (ADC) module. Custom-designed software made fast ADC possible during the whole period of data acquisition. both concepts were applied to a Vickers Medical Mystro electromyograph, and have been succesfully used in the Ljubljana (Slovenia) Institute of Clinical Neurophysiology for the last two years.
motor unit responses; magnetic stimulation; EMG responses. muscle; cortex; brain
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o izdanju
109 (2)
1998.
194-197-x
objavljeno
0924-980X
Povezanost rada
Javno zdravstvo i zdravstvena zaštita