Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Development and validation of a liquid chromatography – tandem mass spectrometry method for the quantification of opiorphin in human saliva (CROSBI ID 175415)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Brkljačić, Lidija ; Sabalić, Maja ; Salarić, Ivan ; Jerić, Ivanka ; Alajbeg, Ivan ; Nemet, Ina Development and validation of a liquid chromatography – tandem mass spectrometry method for the quantification of opiorphin in human saliva // Journal of chromatography. B, 879 (2011), 32; 3920-3926. doi: 10.1016/j.jchromb.2011.11.003

Podaci o odgovornosti

Brkljačić, Lidija ; Sabalić, Maja ; Salarić, Ivan ; Jerić, Ivanka ; Alajbeg, Ivan ; Nemet, Ina

engleski

Development and validation of a liquid chromatography – tandem mass spectrometry method for the quantification of opiorphin in human saliva

Opiorphin, QRFSR-peptide, is a mature product of the PROL1 (proline rich, lacrimal 1) protein that showed beneficial effects in pain management, antidepressant-like actions as well as involvement in colonic motility and erectile physiology. Using opiorphin as a potential biomarker of different pathological states requires the development of robust and sensitive methods. We report a highly sensitive and specific liquid chromatography with tandem mass spectrometric detection (LC– MS/MS) analytical method for the analysis of opiorphin in human saliva. Quantification was based on multiple reaction monitoring using characteristic transitions (m/z 347/120 – as quantifying ion ; 347/175 and 347/268 as qualifying ions). The assay was linear in the range of 0–110 ng/ml and the lower limit of quantification reached was 1.0 ng/ml. The intra-day precision and accuracy were between 2.7–5.6% and −2.3 to 3.2%, respectively. The inter-day precision and accuracy were between 10.8–13.7% and −11.0 to 52%, respectively. Mean recovery was 106% and mean matrix effect was 0.97. Opiorphin in TFA treated saliva samples was stable for at least 12 h at room temperature and up to 30 days at −20 ◦C. Opiorphin levels in human saliva samples collected from young healthy individuals ranged from 2.8 to 25.9 ng/ml.

opiorphin ; saliva ; LC-MS/MS ; biomarker

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

879 (32)

2011.

3920-3926

objavljeno

1570-0232

1873-376X

10.1016/j.jchromb.2011.11.003

Povezanost rada

Kemija, Dentalna medicina

Poveznice
Indeksiranost