Final report on EURAMET.T-K6.1 : Bilateral comparison of the realisations of local dew/frost-point temperature scales in the range −70 °C to +20 °C (CROSBI ID 205411)
Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Heinonen, Martti ; Zvizdić, Davor ; Šestan, Danijel
engleski
Final report on EURAMET.T-K6.1 : Bilateral comparison of the realisations of local dew/frost-point temperature scales in the range −70 °C to +20 °C
As the European extension of the first CCT humidity key comparison, EURAMET.T-K6 was successfully completed in year 2008 [1-2]. After this comparison, a new low dew-point generator was introduced at LPM in Croatia as a result of progress in the EUROMET P912 project. With this new facility, the LPM uncertainties decreased significantly and the operating range became significantly wider. Therefore, it was decided to arrange a bilateral comparison between LPM and MIKES in Finland. As the coordinator of EURAMET.T-K6 and participant of CCT-K6, MIKES provided a link to these comparisons with an appropriate uncertainty level. This comparison was carried out in a manner similar to other K6 comparisons but only one transfer standard was used instead of two units and the measurement point -70 °C was added to the measurement scheme.
Dew-point generator; dew-frost point comparison
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano