A New Sensor for Determination of Anionic Surfactants in Detergent Products with Carbon Nanotubes as Solid Contact (CROSBI ID 239726)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Sakač, Nikola ; Jozanović, Marija ; Karnaš, Maja ; Sak-Bosnar, Milan A New Sensor for Determination of Anionic Surfactants in Detergent Products with Carbon Nanotubes as Solid Contact // Journal of surfactants and detergents, 20 (2017), 4; 881-889. doi: 10.1007/s11743-017-1978-0

Podaci o odgovornosti

Sakač, Nikola ; Jozanović, Marija ; Karnaš, Maja ; Sak-Bosnar, Milan

engleski

A New Sensor for Determination of Anionic Surfactants in Detergent Products with Carbon Nanotubes as Solid Contact

A new solid-state sensor for potentiometric determination of surfactants with a layer of multi-walled carbon nanotubes was prepared. As a sensing material, 1, 3-didecyl-2- methylimidazolium–tetraphenylborate ion-pair was used. The investigated sensor showed a Nernstian response for both dodecylbenzenesulphonate (DBS, 57.6 mV/decade of activity between 5 × 10−7 to 1 × 10−3 M) and sodium lauryl sulfate (LS, 58.4 mV/decade of activity between 2 × 10−7 to 2 × 10−3 M). It responded in 8–10 s for each ten-fold concentration change in the range of 1 × 10−6 to 3 × 10−3 M. The detection limits for DS and DBS were 2 × 10−7 and 3 × 10−7 M, respectively. The sensor revealed a stable response (signal drift 2.6 mV/h) and exhibited satisfactory selectivity performances for LS over most of the anions generally used in surfactant-based commercial detergents. The main application of this sensor was the end-point determination in potentiometric titrations of anionic surfactants. The (diisobutyl phenoxy ethoxy ethyl)dimethyl benzyl ammonium chloride (Hyamine), cetyltrimethylammonium bromide, hexadecylpyridinium chloride monohydrate (HDPC) and 1, 3-didecyl-2-methylimidazolium chloride were tested as potential cationic titrants, and all exhibited analytically usable titration curves with well-defined equivalence points. The standard solution of HDPC was used as a cationic titrant by all potentiometric titrations. The operational life-time of the sensor described was prolonged to more than 3 months.

Surfactant sensor ; MWCNT ; Anionic surfactant ; Potentiometric titration ; Detergents